PRODUCT
首頁

產品資訊

PRODUCT
首頁

產品資訊

Differential AC Amplifier 新款電生理訊號放大器

產品型號:
廠牌名稱:
Model 1700型胞外組織電生理架構簡單且易於操作,非常適合用於常見的體外電生理訊號記錄以及基本的胞外記錄電生理實驗

您可以另外選購記錄探針,或是使用自行設計的探針或是記錄電極

此設備亦可輸出類比訊號,搭配示波器或其他第三方設備配合實驗進行

產品特點

全新改版

  • 接頭改為常見的DB9通用接頭, 方便客戶自行改線
  • 可搭配多種常見電極或是自製電極使用
  • 更佳的機內雜訊遮蔽及長時間使用雜訊量

 

PERFECT FOR EXCITABLE TISSUE RECORDINGS

 

Ideal for multiple-unit recordings, evoked potentials,
EEG/EMG/EKG/ERG recordings, long-term potentiation and more.

 

A-M Systems Model 1700 Differential AC Amplifier

Low Noise

1 µV p-p
Driven Shield Capability

 

Easy to Use

Analog Controls
Record or Stimulate Modes

 

Versatile

Optional Headstages
Customizeable Gain and Controls

 

The Model 1700 Differential AC Amplifier is perfect for excitable tissue recordings, whether in a reseach laboratory or teaching environment.

 

Common applications include, but are not limited to:

Multiple-Unit Recordings

Evoked Potentials

EEG/EMG/EKG/ERG recordings

Long-term Potentiation

 

The Model 1700 Differential AC Amplifier contains 4 independent amplifier channels in a single enclosure enabling the researcher to record multiple signals simultaneously. Each channel contains a high-gain, low-noise differential stage followed by low-pass, high pass, and notch filters. A connection is also provided for a stimulator input so that recording or stimulating through the same electrode is possible without having to change cabling.

 

Standard gain settings are x100, x1000 or x10,000, but they can be modified to meet your specific requirements. High and Low Pass filters are 2nd order Butterworth filters, and like the gain, the standard settings can be modified to meet your needs if required.

 

The Model 1700 Differential AC Amplifier is designed for use with low impedance (<1.5 MΩ) electrodes, and is therefore suitable for use with suction, nerve cuff, hook, surface, and simple metal electrodes. Optional record only headstages are available for situations where you would like to record from electrodes with higher impedances such as those used when recording neural spike action potentials.

產品規格

Noise

  • Voltage, fB = 10 Hz-10 kHz, typical: 1.0 µV, p-p

Offset Voltage

  • Average Drift , maximum: ± 30.0 µV/Deg C

Bias Current

  • Initial Bias Current, maximum: ± 0.4 nA

Offset Current

  • Input Offset Current, maximum: ± 0.4 nA

Input Impedance

  • 1012 Ω || 50 pF

Inter-channel Crosstalk

  • 90 dB at 1 kHz

Voltage Range

  • x100: .11 VAC or .11 VDC ± 5%
  • x1000: .011 VAC or .11 VDC ± 5%
  • x10000: .0011 VAC or .11 VDC ± 5%
  • Common Mode Rejection (CMR): 75 dB

Slew Rate

  • 2 V/µS

Rated Output

  • Voltage Output: ± 11 V
  • Current Output: 5 mA
  • Output Resistance: >5 Ωs

Low Cut-Off Filter

  • Cut-Off Frequencies: 0.1, 1.0, 10.0, 100, 300 Hz
  • Cut-Off Rate: 40 dB/decade

High Cut-Off Filter

  • Cut-Off Frequencies: 0.5, 1.0, 5.0, 10, 20 kHz
  • Cut-Off Rate: 40 dB/decade

Notch Filter

  • Center Frequency: 50 Hz or 60 Hz (factory preset)
  • Rejection: >35 dB