爪蟾卵母細胞電生理研究系統 Xenopus Oocyte Research

產品型號:
廠牌名稱: Multichannel Systems
Multichannel Systems提供了您完整的爪蟾/兩生動物電生理研究解決方案,包含了以下三種商品:
Roboinject 全自動魚卵注射系統
Roboocytes2 全自動魚卵離子通道掃瞄系統
Hiclamp 全自動Voltage-clamp電生理系統

產品特點

 
Roboinject

Fully-automated Roboinject system for automated injection of RNA, DNA or other samples into Xenopus oocytes or cells of similar size. Complete with Roboinject robot, Roboinject software, and accessories.

Key features:

  • Optimized throughput of experiments
  • Automated, high-quality TEVC recording from 96 oocytes without supervision
  • Flexible design of automated recording sequences
  • Automated compound application
  • Automated cell wash
  • Manual recording mode

Read more

Roboocyte2

Fully-automated Roboocyte2 system for screening of ligand-gated and voltage-gated ion channels, as well as electrogenic transporters based on the standard Xenopus oocyte expression system. Complete with Roboocyte2 robot, 12-channel perfusion system, data acquisition and analysis software package, computer, and accessories.

Key features:

  • Optimized throughput of experiments
  • Automated, high-quality TEVC recording from 96 oocytes without supervision
  • Flexible design of automated recording sequences
  • Automated compound application
  • Automated cell wash
  • Manual recording mode

Read more

HiClamp

Automated Voltage-Clamp Screening System for Xenopus laevis oocytes. Complete with HiClamp robot, USB video camera, laptop computer with installed HiClamp software, accessories, and consumables.

Read more

產品規格

Roboinject

Includes:

  • Roboinject robot:
    - Dimensions (W x D x H): 320 mm x 320 mm x 310 mm
    - Supply voltage: 90-264 VAC
    - Connects to the computer via USB interface
    - Injection volume range: 1 - 100 nl
    - Injection precision: 0.5 nl
    - xyz-movement precision: <20 µm
    - Adjustable injection depth
    - Adjustable injection speed
    - Automated sample uptake
  • Roboinject program:
    - Roboinject control program, featuring automated and manual operation
  • Stereo microscope (for alignment procedure)
  • Accessories:
    - Alignment device
    - Oocyte filter
    - Mineral oil
    - Manual
    - 2 spare plunger
  • Consumables:
    - 1 x unit of 10 injection capillaries (IN-25)
    - 20 x 96-well plate from Nunc
    - 250 x 0.5 ml reaction tubes, low retention, DNAse/RNAse free

Technical specifications as specified in the data sheet unless noted otherwise in the order.

Roboocyte2

Includes:

  • Roboocyte2 robot:
    - Dimensions (W x D x H): 320 mm x 320 mm x 310 mm
    - Supply voltage: 90-264 VAC
    - Integrated digital TEVC amplifier, voltage and current clamp operation mode
    - Clamp voltage setpoint: –500 mV to +500 mV
    - Typical rise time in voltage clamp mode: 500 µs
    - Sampling rate: 1 Hz - 20 kHz
    -Current range -105 µA to +105 µA
    - Data resolution: 16 bits
    - Connects to the computer via USB interface
  • Roboocyte2 program:
    - Roboocyte2 control program, featuring automated and manual operation, real-time monitoring, data analysis, database (Microsoft® Access® 2010) linkage
  • Roboflow perfusion system:
    12-channel pinch valve Roboflow-System
    - Two peristaltic pumps
    - Adjustable flow rate 0.1 to 10 ml/min via RS232 serial port
    - Fully controlled and automated by the Roboocyte2 software
  • Computer:
    High end PC with high resolution wide screen monitor, manufactor: DELL®, MCS software installed and tested
    - Operating system (German or English version): Microsoft® Windows® 7 Professional (64bit)
    - Keyboard layout English (other language on request)
    - 8192 MB (2x4096) RAM
    - Min. 1 TB HDD
    - Intel Core I5-2400 (3.1 GHz, 6 MB)
    - CD/DVD Combo drive for backup
    - 24" LED monitor
    - Mouse
  • Stereo microscope (for alignment procedure)
  • Accessories:
    - Manifold and tubing set
    - Alignment device
    - Test model probe
    - Oocyte filter
    - Manuals
  • Consumables:
    - 1 x Unit of 4 TEVC measuring heads
    - 1 x Unit of Microloader tips from Greiner
    - 20 x 96-well plate from Nunc

Technical specifications as specified in the data sheet unless noted otherwise in the order.

References

Documents

 

HiClamp

  • HiClamp:
    - Dimensions (W x D x H): 640 mm x 600 mm x 440 mm
    - Supply voltage: 90-264 VAC
    - Sampling rate of amplifier: 20 kHz
    - Resolution of amplifier: 16 bit
    - Current electrode output: Output range -105 µA to +105 µA
    - Current resolution: 1 nA
    - Voltage range: -100 V to +100 V
    - Voltage electrode input:Input range -500 mV to +500 mV
    - Voltage resolution: 0.015 mV
    - Clamp voltage setpoint range: -500 mV to +500 mV
    - Clamp voltage setpoint resolution: 1 mV
    - Typical rinse time in voltage clamp mode: <1 ms
  • Laptop computer:
    - High end laptop computer, MCS software installed and tested
    - Operating system (German or English version): Microsoft® Windows® 7 Professional (64bit)
    - Keyboard layout English (other language on request)
    - 4 GB RAM
    - 500 GB HDD
    - Intel Core i3-2370 (2.4 GHz, 3 M Cache)
    - 15.6” LED display
  • HiClamp program:
    - HiClamp control program, featuring automated and manual operation
    - Including a 2 years maintenance package (software updates and support)
  • Accessories:
    - Oocyte filter
    - USB video camera
    - Manual
    - 100 micro-stirrers
    • Consumables:
      - 1 x unit of 20 glass electrodes
      - 10 x 96-well plate (conical bottom)
      - 10 x 96-well plate (flat bottom)
      - 2 spare oocyte baskets

    Technical specifications as specified in the data sheet unless noted otherwise in the order.